phoenix v|tome|x c
|
v|tome|x c scatter|correct / HS
|
Minifocus X-ray tube /
|
Closed ISOVOLT 450 M2/0.4-1.0HP
|
Max. voltage / power
|
450 kV @ 700 W/1500 W
|
Focal spot
|
0.4 mm (max. power 700 W) / 1.0 mm (max. power 1500 W)
|
Focus Detector Distance (FDD)
|
1300 mm
|
1150 mm
|
Voxelsize range
|
100 - 146 μm
|
87-139 μm
|
Geometrical magnification (3D)
|
1.37-2x
|
1.44-2.3x
|
Spatial CT resolution
|
2.5 lp/mm at 130 μm voxel resolution referring to ASTM E 1695
|
Detail detectability
|
Down to ~100 μm
|
Measurement accuracy
|
20+L/100 μm referring to VDI 2630-1.3 guideline*
|
Fan beam CT line detector array (LDA)
|
LDA detector package “fan”: 16 bit Linear Detector Array 820 mm sensitive width, 2050 pixels, 400 μm pitch. Linear subpixel-shift axis for resolution improvement and quality enhancement**
|
Cone beam CT DXR flat panel detector (option)
|
Alternative detector package „cone“: 14 bit high-contrast GE DXR 250 detector with 200 μm pixel size, approx. 400 x 400 mm (16“ x 16“) sensitive surface, 2000 x 2000 pixel (4 megapixel). Virtual detector enlargement for large samples
|
Cone beam CT dynamic 41|200 detector (on request)
|
GE dynamic 41|200 large area detector with superior image and result quality, 410 x 410 mm (16” x 16”), 200 .m pixel size, 2036 x 2036 pixels (4 MP), extremely high dynamic range > 10000:1
|
Cone beam CT dynamic 41|100 detector (on request):
|
GE dynamic 41|100 detector 410 x 410 mm (16” x 16”), 100 . m pixel size, 4048 x 4048 pixels (16MP) for doubled CT resolution
|
dual|detector configuration (option)
|
Alternative detector package „cone & fan“: Combination of LDA and flat panel detector, each with detector shift and easy switching between fan beam and cone beam CT mode
|
Granite based high precision Manipulation
|
2 axes (R,Y), manual Z axis (300 mm)
|
3 (R, Y, Z) or 4 (R, Y, Z, XB) in HS base Z-auto- matic or HS quick|pick configuration + detec- tor shift axis
|
Max. 3D scan area d x h / max. weight
|
500 x 1000 mm (270 x 1000 mm scatter|correct) / up to 50 kg (110 lbs.)
|
Max. 3D scan area d x h / max. weight HS
|
|
270 x 310 mm / up to 10 kg (rotation unit HS) 100 x 125 mm / up to 3 kg (quick|pick gripper)
|
Focus object distance
|
650-950 mm
|
500-800 mm
|
System dimensions W x H x D
|
2,310 mm x 2,750 mm x 2,870 mm (91” x 108” x 113”) excluding console
|
System weight
|
~15,000 kg / 33,070 lbs.
|
phoenix datos|x CT software
|
For highly automated one-button|CT including multiple modules for CT data and workflow optimization. Different 3D evaluation software packages for 3D metrology, failure or structure analysis on request
|
phoenix datos|x metrology package (option)
|
surface|extraction – automatic generation of surface data | easy|calib – module for CT system calibration | calibration|object – 1 calibration tool with certificate
|
velo|CT II package (option)
|
For ultrafast volume reconstruction
|
Cabinet crane (option)
|
For ergonomic handling of heavy samples up to 50 kg (110 lbs.)
|
Barcode reader (option)
|
For easy sample identification
|
Radiation protection
|
The radiation safety cabinet is a full protective installation without type approval according to the German RöV. It complies with French NFC 74 100 and the US Performance Standard 21 CFR Subchapter J. For operation, other official licenses may be necessary
|